Standard Equipment

  • Tmax 1200 °C or 1300 °C
  • Single-zoned design
  • Dual shell housing made of textured stainless steel sheets with additional fan cooling for low surface temperature
  • Exclusive use of insulation materials without categorization according to EC Regulation No. 1272/2008 (CLP)
  • Outer tube diameter of 50 mm to 170 mm, heated length from 250 mm to 1000 mm
  • Ceramic working tube C 530 including two fiber plugs for operation under air
  • Thermocouple type N (1200 °C) or type S (1300 °C)
  • Solid state relays provide for lownoise operation
  • Heating elements on support tubes provide for free radiation
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive
  • Controller B510 (5 programs with each 4 segments)

Specifications

Model Tmax1 Outer dimensions3 in mm Outer tube Ø Heated Length constant1 Tube length Connected Electrical Weight
length temperature +/− 5 K in mm load
in °C W2 D H in mm in mm single-zoned three-zoned in mm in kW connection* in kg
R 50/250/12 1200 434 340 508 50 250 80 450 1.6 1-phase 22
R 50/500/12 1200 670 340 508 50 500 170 250 700 2.34 1-phase 34
R 120/500/12 1200 670 410 578 120 500 170 250 700 6.5 3-phase 44
R 170/750/12 1200 920 460 628 170 750 250 375 1070 10.0 3-phase 74
R 170/1000/12 1200 1170 460 628 170 1000 330 500 1400 11.5 3-phase 89
R 50/250/13 1300 434 340 508 50 250 80 450 1.6 1-phase 22
R 50/500/13 1300 670 340 508 50 500 170 250 700 2.34 1-phase 34
R 120/500/13 1300 670 410 578 120 500 170 250 700 6.5 3-phase 44
R 170/750/13 1300 920 460 628 170 750 250 375 1070 10.0 3-phase 74
R 170/1000/13 1300 1170 460 628 170 1000 330 500 1400 11.5 3-phase 89
1Values outside the tube. Difference to temperature inside the tube up to + 50 *Please see page 75 for more information about supply voltage
2Without tube 4Only valid for single-zone version
3External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.